Meriam M2110F SMART FLOW GAUGE User Manual

Page 10

Advertising
background image

10

4.2.4

PRGM or ENTER


Mode of
Operation

PRGM

Document Symbol: (PRGM/Enter

)

Measure Mode

Selects Program Mode, which allows access to the programmable registers P0-P10.

Program Mode

Program Mode (P# displayed): Opens the selected register for editing.

Data Edit Mode: When the desired value is selected, pressing the PRGM/ENTER

key accepts and stores the value.

When editing a multi-digit value, each key press will accept the current digit and proceed
to the next until finally accepting the complete value and closing the register.



4.2.5

ZERO FUNCTION


Mode of
Operation

ENG

UNITS

+

Document Symbol: (Up

and Down)

Measure Mode

In Measure Mode, pressing the Up

and Down↓ keys at the same time resets the zero

reference of the gauge (see page 16).

Program Mode

In Program Mode, for convenience, this function will reset to P0.

Program Mode (P# displayed): Function will reset the register index to P0.

Data Edit Mode: After opening a register for editing, this function will reset the
register’s data value to default.

Note that even when using this “reset” function, the PRGM/Enter

→ key must be

pressed to accept the new value. This allows for further editing after reset of the value.


Advertising