13 SUPPLEMENTARY INFORMATION


13.1 Application Example

13.1.1 Laminar flow element

Primary Element:

Laminar Flow Element (LFE)

Model

50MJ10-9

Nominal Air flow Range = 3.00 cubic feet/minute

B-Coefficient

=

4.2847E-1

C-Coefficient = -5.0523E-3


Gauge requirement:

M2110F with 10.00 inH2O sensor


Application requirements:

Monitor airflow in units of cubic feet/minute

Setup Relay outputs to detect low limit at 1.25 cu.ft/min


Settings:

P3 = 1

: Measurement Model = LFE

P5 = 1

: Setpoint Option = SET1 Only

P6 = 1.250

: SET1 limit = 1.250 cu.ft/min

P9 = 4.285

: Rounded decimal part of the B-Coefficient

= - E1

: Exponent of B-Coefficient

P10 = -5.052

: Rounded decimal part of the C-Coefficient

= - E3

: Exponent of C-Coefficient

13.1.2 Accutube/Orifice Plate flow element

Primary Element:

Accutube Flow Element

Maximum

Air

flow

Range

=

750

SCFM

DP at maximum air flow = 15 inH

2

O


Application requirements:

Monitor airflow in units of cubic feet/minute

Setup Relay outputs to detect low limit at 25 cu.ft/min


Gauge requirement:

M2110F with 28.000 inH

2

O sensor


Settings:

P3 = 2

: Measurement Model = Square Root Device

P5 = 1

: Setpoint Option = SET1 Only

P6 = 25

: SET1 limit = 25 cu.ft/min

P9 = 193.65

: Flow Constant =

mAirFlow

DPatMaximu

FlowRange

MaximumAir


13.2 Maintenance

13.2.1 Battery Model Smart Gauge

The battery can be changed by removing the front face cover on the smart gauge and then removing the PC board
assemblies as described below.

1. Remove the four (4) front face screws.
2. Carefully fold down front face cover and slide electronics board assembly forward until the sensor ribbon cable

can be unplugged.

3. Slide the assembly completely out of the housing, unplug battery board and attach replacement.
4. Re-assemble the unit in reverse order.
5. Check to ensure that the sensor cable, battery board and front panel are re-connected properly before reattaching

the front faceplate.