Range limits and indication – Meriam M2110F SMART FLOW GAUGE User Manual

Page 35

Advertising
background image

35


Range Limits and Indication


Sensor Hard

Soft

Certified

Soft

Hard

Type

Under Under

Range

Over

Over

Range Range

(F.S.)

Range

Range

DN

Under

-20%

1

-20 to 0%

0 to 100%

100 to

120%

Over

120%

D

I

Under

-20%

2

-20 to 0%

0 to 100%

100 to

120%

Over

120%

G

I

Under

-20%

3

-20

3

to

0%

0 to 100%

100 to

120%

Over

120%

C

I

NA

4

NA

4

0 to 100%

100 to

120%

Over

120%

A

I

NA

5

NA

5

0 to 100%

100 to

120%

Over

120%

Notes:

1

DN units will measure to -20% (or -10 PSID, whichever is greater

pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).

2

D

I

units will measure to -20% (or -150 PSI, whichever is greater

pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).

3

G

I

units will measure to -20% (or -10 PSIG, whichever is greater

pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).

4

C

I

units will measure to negative barometric pressure (typically -14.7

PSIG) and to +120% of full scale pressure. Calibration certification is for
-14.5 PSIG to F.S. (Full Scale Only).

5

A

I

units will measure to +120% of full scale pressure. Calibration

certification is for 0 PSIA to F.S. (Full Scale Only).

Hard Over & Under:
Display will read "OP" (Under -20% or Over 120%).

Soft Over & Under:
Display will still read out at (± 20%).

Advertising