Section G: Boards

14

SW3

Function

OFF

ON

1

Heated Line Control

OFF

ON

2

Probe Control

OFF

ON

3

Peltiers

1 per channel

2 per channel

4

Channels

1

2

5

Peltier Power Share

Full Power

Power Sharing

6

ISO Channels

Inactive

Displayed


SW2

Function

ON

OFF

1

ICE VCC

Enabled

Disabled

2

JTAG ICE ResetN

Enabled

Disabled

3

Calibration

Enabled

Normal Operation

4

Dflash Res

Enabled

Normal Operation

5

Init Clr

Enabled

Normal Operation

Button Control

The buttons are used to control the calibration settings. Follow the bottom line on
the display to determine the functions in the different modes. The three
combinations that are available are left only, right only, or both. To select the center
option on the bottom line of the display, press both buttons together.

Calibration Menus

The calibration mode allows the board to be setup with min and span points as well
as all of the other configuration information applicable depending on the
configuration switch settings.

Ch1, Ch2, TC1, TC2
Min = 0
Span = 5 degrees C
Max = 8*Span Temp (360°C limit)

The calibration is performed by first calculating the expected settings for the
calibration resistors before entering the calibration operation to get a good starting
point. Then the 0 point is adjusted to the minimum point in the ADCC (memory
chip). The span temperature is then set and the span gain is adjusted to make the
span point 1/8 of the entire range for the temperature measurements.

ChX: MIN = xx

C

ChX: SPAN = xx

C