Section G: Boards

15

ChX:RDY TMP= xx

C

ChX:SET TMP= xx

C

ChX GAIN=%3u


Probe, Heated Line
Min = 0°C
Span =
Max = 360°C

Calibration is performed by setting the gain of the span circuit to minimal gain then
adjusting the 0 point to the minimum reading from the ADCC. The span temperature
is used to measure the ADCC value and calculate the step size of the ADCC.

ChX: MIN = xx

C

ChX: SPAN = xx

C

ChX:MIN TMP= xx

C

ChX:MAX TMP= xx

C

ChX:SET TMP= xx

C

ChX: GAIN=%3u


Isolation Channels
Min = 0°C
Max = 25°C

The temperature range for the isolation circuit is 0 to 25°C. The calibration is
performed by first calculating a good starting point for the offset and span resistors.
Then adjust the 0 point to the minimum point in the ADCC. The last step is to set
the input to 25°C and adjust the span gain to achieve the max value in the ADCC.

ChX: MIN = xx

C

ChX: MAX = xx

C

ChX:MIN TMP= xx

C

ChX:MAX TMP= xx

C

ChX:SET TMP= xx

C

ChX: GAIN=%3u

Water Slip (Moisture Carryover) Relay Board

Note: For the following instructions, typical applications run each of the relays in
series through the Common and Normally Open (NO) contacts.

A: Computer Status Alarms (External PLC monitoring).

Wire computer status alarms to JP4 & JP2. Wire the Common and Normally
Open Contacts in series to a PLC. If an alarm occurs, one or both of these
contacts will open.