Section H: Default Setpoints

17

H:

D

EFAULT

S

ETPOINTS


eCOOL 5000™ Heat Exchangers:
Setpoint Temperature = 5°C
Run Temperature = 10°C
Cooler must operate below the Run Temperature to avoid the over temperature
alarm condition.

Heated Sample Probe:
Setpoint Temperature = 190°C
High Temperature Alarm = 210°C
Low Temperature Alarm = 120°C

Heated Sample Line:
Setpoint Temperature = 190°C
High Temperature Alarm = 210°C
Low Temperature Alarm = 120°C

LAN Card:
Static IP Addressing: 192.168.45.140
DHCP Disabled.

Relay Card:
Relay contacts are in their Normal state (de-energized state) during an alarm
condition.

Analog output 0-4.5 VDC equates to 0-10°C corresponding to each thermocouple.