A: s, Pecifications, General specifications – Perma Pure Baldwin-Series MP5400 User Manual
Page 3: Digital board specifications, Operating specifications
Section A: Specifications
3
A: S
PECIFICATIONS
General Description
Digital control board
Dual channel system
2 x 5 inch heat exchangers connected in series (single stream) or parallel (dual stream)
Jumbo 2-line LCD temperature and alarm display
Single head mini sample pump
Dual head mini peristaltic drain pump
Model Specifications
Model Standard
Capacity
Heat Exchangers
Passive Active
Dimensions Weight
Power
Supply
MP5400
3-4 LPM
6-8 SCFH
---
2x5 in.
10 x 9 x 18 in. HWD
25 x 23 x 46 cm HWD
27 lbs
12 kg
240W
General Specifications
Digital Boards
Main control board
Water slip alarm relay board
Alarms
Cooler over temperature
Cooler thermocouple failure
Water slip (optional)
Display
Jumbo dual-line LCD
LED red, amber, green heat exchanger status
indicators
Heat Exchanger
Type
EZ-Clean™ twist apart
Heat Exchanger
Material
Stainless steel (standard); Durinert
®
treated
stainless steel, Kynar
®
(optional)
Heat Exchanger
Connections
1/4” FNTP Inlet
1/8” FNTP Outlet
3/8” FNTP Drain
Heat Sink
High heat transfer aluminum
Voltage
110/120 VAC, 220/240 VAC, 12 VDC, 15 VDC,
24 VDC
Digital Board Specifications
Main Control Board
4 thermocouple inputs
4 analog outputs
Water Slip Alarm Relay
Board
2 water slip (moisture carryover) inputs
2 high current digital contact outputs
2 digital PLC outputs
Operating Specifications
Maximum Inlet Sample
Temperature
400
°F (205°C) SS, Durinert
®
, Glass Impingers
280
°F (138°C) Kynar
®
Impingers
Maximum Inlet Pressure
45 psig
3 bar / 2250 mmHg
Maximum Heat Exchanger
Pressure Drop
<+1 in. H2O
Ambient Temperature
Range
33-104
°F
0.6-40
°C
Outlet Sample Gas Dew
Point
41
°F
5
°C