Section E: Start-up Procedure

8

E: S

TART

-

UP

P

ROCEDURE


Plug in the power cord to a properly grounded main circuit. The LCD will display the
actual temperature. The temperature of each heat exchanger will fall until it reaches
approximately 5°C.

If there are no alarms (an alarm condition consists of water slip detection after the
cooler, thermocouple failure, or actual temperature > 10°C), then the LED(s) should
turn to an amber color. This indicates that the alarms are cleared and the user can
PRESS THE RESET BUTTON, located on the back of the unit, to begin operation.
Once the reset button is pressed, the LED(s) should turn green indicating that there
are no alarms and sample flow can begin. If the sample pump is wired through the
relay board alarm contacts and all alarms are cleared, the sample pump will start.
The sample pump will stop if any alarms are detected.

Note: Current alarm conditions are displayed on the second line of the LCD.

The eCOOL™ Mini Digital Portable Gas Sample Conditioning System is virtually
maintenance free. However, in the event of electrical problems, refer to the
Troubleshooting section in this manual.