Operations – Teledyne TED-200 - Medical application oxygen monitor User Manual

Page 11

Advertising
background image

Operations 2

2-1

Teledyne Analytical Instruments

Operations

Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the unit

and all included accessories for broken or loose parts. If damaged, DO
NOT USE. Notify the shipper, and consult Teledyne Electronic Technolo-
gies, Sensor Technologies.

Note: This equipment is internally powered.

2.1

Setup

The TED 200-T is suitable for use in many medical applications. The unit can be

used on a table top or mounted on a pole using a mounting bracket (P/N B34102).

To set up and use your TED 200-T:

1. Install the batteries.

2. Install the sensor.

3. Calibrate the unit.

4. Set the alarms.

The front display touch panel is designed to insure that accidental touching or

bumping of the panel will not reset any calibrations or settings. Each button used in a
sequence must be pressed within five seconds of each other, and the CAL, ALARM
TEST and OFF keys must be pressed twice in order to activate them. The ON/OFF
key needs to be pressed once in order to turn on the TED 200-T, but must be pressed
twice to turn it off.

Advertising