Teledyne TED-200 - Medical application oxygen monitor User Manual

Page 13

Advertising
background image

Operations 2

2-3

Teledyne Analytical Instruments

Figure 3: Mounting the sensor in the tee adapter

6. Close the compartment door. The front panel will now display HI ALARM

100 and LO ALARM 17 with a flashing AIR CAL.

7. To test the batteries after the unit has been calibrated: Press the BATT

TEST key once. The display should show BATTERY HRS LEFT as 999.

2.1.2 Sensor Installation or Replacement

Note: The T-7 oxygen sensor must be installed before the TED 200-T can be

operated. For sterilization procedures, see page 2.8: Sterilization.

Note: When a new sensor is installed, typically, the unit is suitable for use

immediately after installation.

CAUTION:

Do not autoclave the sensor.

1. Remove the new sensor from its protective bag. Inspect the sensor for

damage or electrolyte leakage. If the sensor is damaged, obtain a
replacement. Do not use the defective sensor as it may damage the unit.

WARNING: THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT LET IT

COME IN CONTACT WITH SKIN. IF IT DOES, FLUSH AF-
FECTED AREA WITH WATER. DO NOT ATTEMPT TO OPEN
OR REPAIR THE SENSOR. LEAKING OR EXHAUSTED SEN-
SORS SHOULD BE DISPOSED OF IN ACCORDANCE WITH
LOCAL REGULATIONS. CONSULT THE MATERIAL SAFETY
DATA SHEET IN THE APPENDIX.