Low flow applications (5 to 500 ml/min), Setup – SKC 224-PCXR4 Universal Pump User Manual

Page 11

Advertising
background image

9

Low Flow Applications (5 to 500 ml/min)

Using Single Adjustable Low Flow Holder

Figure 2

Front, back, and top views of PCXR4 Sampler

For additional drawings, see pages 21 and 23.

Setup

Install battery (see Installing the Battery Pack on page
18
). For optimum charge, ensure pump is not running.
Charge the battery by connecting the charger plug to
the sampler charging jack (Figure 2, #18). Ensure that
the battery is fully charged before sampling.

After charging the battery pack, it is good practice

to run the pump for approximately fi ve minutes

before calibrating. This ensures the battery is in

more steady-state conditions and improves the

agreement in pre and post-sampling calibrations.
Do not charge or operate pump from charger in hazardous locations.
Use only an SKC-approved charger designated for this model to ensure

reliable performance. Failure to do so voids any warranty.

Ensure proper orientation of charging cable before plugging it into

the charging jack. Improper orientation/contact will short-circuit the

battery and voids any warranty.

Short-circuiting the battery pack will render it immediately inoperative.

Failure to follow warnings and cautions voids any warranty.

The battery pack may be kept on the SKC-approved charger for an

indefi nite time.

AIRCHEK

SAMPLER

5

4

3

2

1

SAMPLE PERIOD Ñ MINUTES

START

HOLD

FLOW

AND

BATTERY

CHECK

FLOW

ADJ

ON

AIRCHEK SAMPLER

MODEL 224-PCXR4

LISTED 124U

4

3

2

1

BACK

TOP

FRONT

5

16

16

18

8

9

9

14

7

15

Charger and battery

connected

1 LCD

2 Flow and Battery

Check Key

3 Start/Hold Key

4 On/Off Switch

5 Tamper-resistant Cover

7 Flow Adjustment Screw

8 Accessory Mounting

Screws (2)

9 Intake/Filter Housing

14 Cap Screw to Regulator

16 Battery Pack Screws (2)

18 Charging Jack

1