SKC 224-PCXR4 Universal Pump User Manual

Page 12

Advertising
background image

10

Activating the Regulator

Remove the tamper-resistant cover.
Start the pump using the on/off switch
(Figure 2, #4). Press Start/Hold (Figure
2, #3). Press Flow and Battery Check
(Figure 2, #2). Adjust fl ow using the
fl ow adjustment screw (Figure 2,
#7) until the built-in rotameter reads
1.5 L/min. The LCD should indicate
BATT OK in the upper left corner (if
it doesn’t, recharge the battery). Press
Flow and Battery Check to place the
pump in Hold.

Remove the cap screw covering the regulator valve (Figure 2, #14)
and turn the exposed screw four to fi ve turns
counterclockwise.

Replace the cap screw. The pump is now set for low fl ow.

For low fl ow,

turn valve screw

counterclockwise.