SKC 224-PCXR4 Universal Pump User Manual

Page 13

Advertising
background image

11

Setting or Verifying Flow Rate

For a diagram of the pump, see Figure 1, page 5.

Before use, allow pump to equilibrate after moving it from one

temperature extreme to another.

Ensure pump has run for fi ve minutes before proceeding with
calibration.
Connect a single adjustable low fl ow holder (Figure 3) to the pump
intake (Figure 2, #9) using 1/4-inch Tygon tubing.
Insert an opened sorbent tube (Figure 3, #3) into the rubber sleeve
(Figure 3, #2) of the low fl ow holder with the arrow on the tube
pointing toward the holder.
Connect a calibrator to the exposed end of the sorbent tube.
Loosen the brass fl ow adjust screw on
the low fl ow holder. Activate the pump
by pressing Flow and Battery Check.
Adjust the fl ow rate by turning the
fl ow adjust screw (Figure 3, #1) on the
holder until the calibrator indicates the
desired fl ow.

Do not adjust the fl ow on the

pump. Adjust the fl ow only by

using the fl ow adjust screw on
the low fl ow holder.

When the desired fl ow is set, place
pump in Hold by pressing Flow and
Battery Check. Disconnect the calibrator.

Figure 3

Single Adjustable
Low Flow Holder

with sample tube

1 Flow adjust screw

2 Rubber sleeve

3 Sorbent tube

Air

fl ow

Calibration train with

tube in low fl ow holder

Flow adjust

screw

Turn screw on

low fl ow holder

to adjust fl ow.

continued on page 12