15

Turn the fl ow adjust screw (Figure 4, #2) for the appropriate
port of the low fl ow holder until the desired fl ow rate is
achieved through the tube. Turn clockwise to decrease the
fl ow.

Do not adjust the fl ow on the pump. Adjust the fl ow only by

using the fl ow adjust screw on the low fl ow holder.

Do not exceed 500 ml/min fl ow rate per tube.

When the desired fl ow is set on the
initial tube, place pump in Hold by
pressing Flow and Battery Check.
Remove the calibrator from the tube
and connect to the exposed end of
the next sorbent tube. Press Flow and
Battery Check and repeat the fl ow
adjustment process until all tubes are
fl ow calibrated. Changing the fl ow on
one tube will not affect the fl ow rate
through the remaining tubes.

Do not exceed 500 ml/min fl ow rate per tube.

For tri and quad models, fi rst rotate each anti-tamper cover
(Figures 4 [on page 13] and 5) to expose the fl ow adjust
screws, then adjust the appropriate screw until calibrator
indicates the desired fl ow.

When the fl ow rate is set for each tube, press Flow and
Battery Check to place the pump in Hold and disconnect the
calibrator.

Replace the sampling media used for calibration with
unexposed media for sample collection. Use a protective tube
cover to prevent tube breakage.

If sampling with fewer tubes than number of ports, insert

unopened sorbent tubes in the empty ports to seal them.

Screw

Screw

Figure 5 -

Cut-away of Tri/Quad Low

Flow Holder

2

cont’d