SKC 224-PCXR8 Universal Pump User Manual

Page 14

Advertising
background image

12

Setting or Verifying Flow Rate

For a diagram of the pump, see Figure 1, page 5.

Before use, allow pump to equilibrate after moving it from one temperature
extreme to another.

Ensure pump has run for fi ve minutes
before proceeding with calibration.

Connect a single adjustable low fl ow
holder (Figure 2) to the pump intake
(Figure 1, #13) using 1/4-inch Tygon
tubing.

Insert an opened sorbent tube (Figure 2, #3) into the rubber
sleeve (Figure 2, #2) of the low fl ow holder with the arrow on the
tube pointing toward the holder.

Connect a calibrator to the exposed end
of the sorbent tube.

Loosen the brass fl ow adjust screw on
the low fl ow holder. Activate the pump by
pressing Flow and Battery Check.

Adjust the fl ow rate by turning the fl ow adjust
screw (Figure 2, #1) on the holder until the
calibrator indicates the desired fl ow.

Do not adjust the fl ow on the pump. Adjust the fl ow only by using the
fl ow adjust screw on the low fl ow holder.

Calibration train with tube

in low fl ow holder

Figure 2

Single Adjustable
Low Flow Holder

with sample tube

1 Flow adjust screw

2 Rubber sleeve

3 Sorbent tube

Air

fl ow

3

Turn screw on

low fl ow holder to

adjust fl ow.

Flow adjust

screw

continued on page 13