Setting or Verifying Flow Rate

When performing multiple-tube sampling using an adjustable low fl ow
holder (dual, tri, or quad), ensure the regulator has been activated and the
pump fl ow rate is set at 1.5 L/min. The maximum fl ow rate through any one
tube is 500 ml/min*. Calculate the sum of all tube fl ow rates. If the sum is
≤ 1000 ml/min, proceed with calibration and sampling without any further
adjustment to pump fl ow rate. If the sum is > 1000 ml/min, set the pump
fl ow rate 15% higher than the sum of tube fl ow rates.

* Back pressure across some sample tubes can be higher than average. In these instances,

the maximum fl ow rate of 500 ml/min per tube may not be achieved.

Before use, allow pump to equilibrate after moving it from one temperature
extreme to another.

Ensure pump has run for fi ve minutes before proceeding with
calibration.

Ensure the pump is set for low fl ow (see Activating the Regulator,
page 11)
.

Connect the adjustable low fl ow holder (Figure 3, page 15) to the
pump intake (Figure 1, #13) using 1/4-inch Tygon tubing.

Insert an opened sorbent tube into each rubber sleeve of the
low fl ow holder (Figure 3, #3 and 4) with the arrow on the tube
pointing toward the holder.

If sampling with fewer tubes than number of ports, insert unopened
sorbent tubes in the empty ports to seal them.

Note the fl ow rates specifi ed by each sampling method and add
them together. If the sum is ≤ 1000 ml/min, proceed to the next
step. If the sum is > 1000 ml/min, multiply the total tube fl ow rate
by 1.15 and set the pump for that fl ow rate.

Connect a calibrator to the exposed end of a sorbent tube, loosen
the brass fl ow adjust screw on the low fl ow holder, and activate
the pump by pressing Flow and Battery Check.

Connect holder to

pump intake and tube

inlet to calibrator.