Preparation and assembly, Handling, Cleaning the sampler – SKC 225-70A IOM Inhalable Dust Sampler with MultiDust Foam Discs User Manual

Page 3

Advertising
background image

3

Preparation and Assembly

Handling

Wear powder-free gloves when handling cassett es and use forceps when working
with fi lters and foams to prevent the transfer of moisture, dust, or other contami-
nants onto the sampling media. Users may wish to use sterile gloves and forceps
depending on the contaminant of interest.

Cleaning the Sampler

1. Disassemble the IOM Sampler (Figures 1, 2, and 3).
2. Place parts in an ultrasonic cleaner with water and a wett ing agent such as a mild

soap. IOM components may also be cleaned with a solvent such as isopropyl
alcohol. O-rings should be cleaned separately using water only.

Do not use CFC-based substances to clean the plastic IOM. Do
not use solvents on O-rings.

When using solvents, protect against splashes to the eye, vapor
inhalation, and vapor ignition.

3. Clean the components using a clean lint-free paper, cloth, or soft brush. Allow

components to dry completely.

Sterilizing Filter, Foam Disc, and Cassette (for Bioaerosol Sampling)

Polycarbonate Filter
Autoclave polycarbonate fi lter according to manufacturer instructions.

MultiDust Foam Disc
1. Place foam in a sonic bath of distilled water for 20 minutes. Remove from

bath.

2. Rinse foam with distilled water.
3. Dry overnight at 158 F (70 C). Equilibrate to standard conditions of

temperature and humidity.

Do not desiccate foam.

4. Expose foam to ultraviolet light according to a standard UV light exposure

method.

Proper O-ring placement in

IOM front plate, plastic and

stainless steel models

Proper O-ring placement in housing

body, plastic model only

O-ring A

O-ring B

O-ring C

Figure 3. O-ring placement