Transporting loaded cassettes, Loading a cassette into the iom housing body – SKC 225-70A IOM Inhalable Dust Sampler with MultiDust Foam Discs User Manual

Page 6

Advertising
background image

6

5. Place a single MultiDust Foam Disc for

respirable sampling (Cat. No. 225-772) into
the inlet of the cassett e top so that it sits just
below the rim (Figure 7). Ensure that it is
not creased and there are no gaps between
it and the cassett e wall.

Note: The foam disc is larger than the
inlet; therefore, it must be compressed
evenly to fi t properly.

6. Insert the cassett e top containing the foam

disc into the cassett e bott om ensuring the
fi lter is in place (see Filter Cassett e: Opening,
Installing Filter, and Closing
).

7. Re-weigh the entire assembly and record pre-weight #2. If the cassett e cover is

included in this operation, reference it to the cassett e with a number or lett er. It
is recommended that the cassett e cover remain in place until ready to sample.

Transporting Loaded Cassettes

If not ready to sample immediately or if transport-
ing pre-weighed loaded cassett es to a sampling
location, place the cassett e with cover into the
provided transport clip (Figure 8). Ensure the
cassett e cover is in place. An alternative is to insert
the loaded cassett e into an IOM housing body and
place the cover on the cassett e opening (See Load-
ing a Cassett e into the IOM Housing Body and Figure
9
). Users may wish to transport loaded cassett es
or loaded IOM housing bodies in a sterile plastic
bag depending on the contaminant of interest.

Loading a Cassette into the IOM Housing Body

Reminder: Use powder-free gloves and forceps to handle cassett e, fi lter, and foams.
Users may wish to use sterile gloves and forceps depending on the contaminant
of interest.

1. Unscrew front plate from IOM housing body.
2. Ensure the O-rings are positioned correctly (Figures 1, 2, and 3).
3. Insert the cassett e into the IOM housing body.
4. Screw the front plate into the housing body. Tighten securely to achieve a good

seal. Install cover on inlet until ready to sample.

Figure 9. Load pre-weighed

cassette into an IOM housing body;

include cover.

Cover

Front

plate

Cassette

assembled

with fi lter

Housing

body

Figure 8. Insert cassette with cover

into transport clip

Foam

insert

Cassette

front

Filter

Cassette

support

grid

Figure 7. Foam inserted into cassette