Transporting Samples to a Laboratory

Wear powder-free gloves when handling cassett es and use forceps
when working with fi lters. Users may wish to use sterile gloves
and forceps depending on the contaminant of interest.

For gravimetric samples:

Immediately following sampling, unscrew IOM front plate and remove cassett e
from housing body. Place cassett e cover on cassett e and insert into a transport clip
(Figure 8). Place unit into a sterile plastic bag depending on the contaminant of in-
terest. Package in a padded envelope and send to a laboratory for analysis. Include
pertinent sampling information and blanks (see Sample Blanks). If post-weighing on
the premises, see Analysis.

For chemical samples (stainless steel cassette only):

1. Unscrew the IOM front plate.
2. Remove the stainless steel cassett e.
3. Choose from the following options:

a. Immediately place entire cassett e/fi lter assembly into a large vial fi lled

with method-specifi ed desorbing solution and cap the vial or,

b. Immediately remove cassett e top and use forceps to remove fi lter. Place

fi lter in a vial fi lled with method-specifi ed desorbing solution and cap
the vial.

4. Package securely to prevent leakage and breakage, include pertinent sampling

information and blanks, and send to a laboratory for analysis.

For bioaerosol samples:

Transport samples in vials fi lled with Peptone Inositol Tween (PIT), a biological
buff er, as indicated in Method 1 or 2 on page 10. Users may wish to place vials in a
sterile plastic bag depending on the contaminant of interest. Samples can be stored
overnight at 39.2 F (4 C) in vials or bags and analyzed the next day.

Sample Blanks

Load and handle the blank IOM Sampler in the same manner as the IOM used for
sampling. Do not pull air through the blank. Send the blank with the samples to
a laboratory.

Analysis

Post-weighing for Inhalable Fraction

Wear powder-free gloves when handling cassett es and use forceps
when working with fi lters.

1. Equilibrate fi lter and plastic cassett e for a reasonable time based on humidity

conditions.

2. Remove the cassett e from the transport clip or sampler housing body. The cassett e

cover can remain on the cassett e if it was included in the pre-weight.

3. Weigh the entire cassett e. Compare the pre and post-weights. The diff erence

between weights is the amount of inhalable particles collected.

The IOM cassett e can be cleaned, reloaded with a fresh fi lter, and reused. See Clean-
ing the Sampler on page 3
.