IOM Stainless Steel Cassett e
Use stainless steel cassett e only.
Autoclave stainless steel cassett e up to 392 F (200 C).

Reminder: Use powder-free gloves and forceps to insert the fi lter and foam into the
cassett e (see Filter Cassett e: Opening, Installing Filter, and Closing). Users may wish to
use sterile gloves and forceps depending on the contaminant of interest.

Storage Requirements for IOM MultiDust Foam Discs

IOM MultiDust Foam Discs are sensitive to heat, humidity, and ultraviolet light.
The foams must be stored in the supplied packaging (including the resealable bag).
Keep in a cool, dry place away from direct sunlight. SKC recommends checking the
condition of the foams before use. Note that the expiration date on the package is
subject to correct storage as detailed above.

General Weighing Guidelines

• The plastic cassett e with fi lter and MultiDust Foam Disc (if used) should be

equilibrated overnight in a balance room under controlled humidity conditions
and weighed. Maintain a stable humidity level in the balance room. A stainless
steel cassett e with fi lter may be desiccated.

• Before pre and post-weighing, wipe the external surface of the sampler with a

clean lint-free paper, cloth, or soft brush.

• For inhalable only, pre and post-weigh cassett e with fi lter as a single unit. For

inhalable and respirable, pre and post-weigh fi lter, foam, and cassett e together
fi rst for inhalable result and then fi lter and cassett e bott om together for respirable
result.

• Field blanks can be used to correct weights when using plastic cassett es at low

fi lter loadings. See Sample Blanks on page 9.

The IOM cassett e assembled with fi lter or the fi lter alone should be weighed

on a fi ve-fi gure forced balance. Allow the cassett e three-and-a-half minutes to
stabilize before taking a reading. The same balance should be used for both pre
and post-weighing procedures.

Pre and Post-weighing for Specifi c Fractions

Particle Size Collected

Pre-weigh Components

Post-weigh Components

Inhalable only

Filter and cassette together

Filter and cassette together

Respirable (#1) and

Inhalable (#2)

#1 Filter and cassette

bottom together

#2 Filter, foam, and cassette

together

#1 Filter and cassette

bottom together

#2 Filter, foam, and cassette

together