Performance characteristics, Continued) – Agilent Technologies 54600-Series User Manual

Page 14

Advertising
background image

14

Performance Characteristics

(continued)

Vertical System: Analog Channels (continued)

Range

1,2

54621A/D, 54622A/D, 54624A: 1 mV/div to 5 V/div
54641A/D, 54642A/D: 2 mV/div to 5 V/div

Maximum Input

CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50

input

Offset Range

54621A/D, 54622A/D, 54624A:

±5 V on ranges <10 mV/div; ±25 V on ranges 10 mV/div to 199 mV/div; ±100 V on ranges

200 mV/div

54641A/D, 54642A/D:

±5 V on ranges <10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div; ±75 V on ranges >200 mV/div

Dynamic Range

Lesser of ±8 div or ±32 V from center screen

Input Resistance

54621A/D, 54622A/D, 54624A: 1 M

±1%

54641A/D, 54642A/D: 1 M

±1%, 50

selectable

Input Capacitance

54621A/D, 54622A/D, 54624A: ~ 14 pF
54641A/D, 54642A/D: ~ 13 pF

Coupling

54621A/D, 54622A/D, 54624A: ac, dc, ground
54641A/D, 54642A/D: ac, dc

BW Limit

54621A/D, 54622A/D, 54624A: ~ 20 MHz selectable
54641A/D, 54642A/D: ~ 25 MHz selectable

Channel-to-Channel Isolation

54621A/D, 54622A/D, 54624A: dc to 20 MHz > 40 dB; 20 MHz to max bandwidth > 30 dB

(with channels at same V/div)

54641A/D, 54642A/D:DC to max bandwidth >40 dB

Probes

54621A/D, 54622A/D, 54624A: 10:1 10074C shipped standard for each analog channel
54641A/D, 54642A/D: 10:1 10073C shipped standard for each analog channel

Probe ID (Agilent/HP and

Auto probe sense

Tek compatible)

ESD Tolerance

±2 kV

Noise Peak-to-Peak

54621A/D, 54622A/D, 54624A: 2% full scale or 1 mV, whichever is greater
54641A/D, 54642A/D: 3% full scale or 3 mV, whichever is greater

Common Mode Rejection Ratio

20 dB @ 50 MHz

DC Vertical Gain Accuracy*

1,2

±2.0% full scale

DC Vertical Offset Accuracy

54621A/D, 54622A/D, 54624A:

<200 mV/div: ±0.1 div ±1.0 mV ±0.5% offset value;

200 mV/div: ±0.1 div ±1.0 mV ±1.5% offset value

54641A/D, 54642A/D:

200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value; >200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value

Single Cursor Accuracy

1,2

±{DC Vertical Gain Accuracy + DC Vertical Offset Accuracy + 0.2% full scale (~1/2 LSB) }
54621A/D, 54622A/D, 54624A example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,

accuracy = ±{2.0%(80mV) + 0.1 (10 mV) + 1.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 3.78 mV

Dual Cursor Accuracy*

1,2

±{DC Vertical Gain Accuracy + 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,

accuracy = ±{2.0%(80 mV) + 0.4%(80 mV)} = ±1.92 mV

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

1

For 54621A, 54621D, 54622A, 54622D, and 54624A, 1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for
1 mV/div sensitivity setting.

2

For 54641A, 54641D, 54642A, and 54642D, 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.