I

2

C

Trigger on I

2

C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data

values. Also trigger on Missing Acknowledge, Restart, EEPROM read, and 10 bit write.

SPI

Trigger on SPI (Serial Protocol Interface) data pattern during a specific framing period. Support positive and negative Chip
Select framing as well as clock Idle framing and user specified number of bits per frame.

Duration

Trigger on a multi-channel pattern whose time duration is less than a value, greater than a value, greater than a time value
with a timeout value, or inside or outside of a set of time values.
Minimum duration setting: 5 ns
Maximum duration setting: 10 s

Autoscale

Finds and displays all active analog and digital (for 54621D/54622D/54641D/54642D) channels, sets edge trigger mode on
highest numbered channel, sets vertical sensitivity on analog channels and thresholds on digital channels, time base to
display ~1.8 periods. Requires minimum voltage >10 mVpp, 0.5% duty cycle and minimum frequency >50Hz.

Analog Channel Triggering

Range (internal)

±6 div from center screen

Sensitivity*

54621A/D, 54622A/D, 54624A: Greater of 0.35 div or 2.5 mV
54641A/D, 54642A/D:<10mV/div: greater of 1 div or 5mV;

10mV/div: 0.6 div

Coupling

AC (~3.5 Hz on 54621A/D, 54622A/D, 54624A.
~10 Hz on 54641A/D, 54642A/D), DC, noise reject, HF reject and LF reject (~ 50 kHz)

Digital (D15 - D0) Channel Triggering (54621D, 54622D, 54641D, and 54642D)

Threshold Range (used defined)

±8.0 V in 10 mV increments

Threshold Accuracy*

±(100 mV + 3% of threshold setting)

Predefined Thresholds

TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V

External (EXT) Triggering

Input Resistance

54621A/D, 54622A/D, 54624A: 1 M

, ±3%

54641A/D, 54642A/D: 1 M

±3% or 50

Input Capacitance

54621A/D, 54622A/D, 54624A: ~ 14 pF
54641A/D, 54642A/D: ~13pF

Maximum Input

CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-ohm input

Range

54621A/D, 54622A/D, 54624A: ±10 V
54641A/D, 54642A/D: DC coupling: trigger level ± 8V; AC coup./LFR: AC input minus trig level not to exceed ±8V

Sensitivity

54621A/D, 54622A/D, 54624A: dc to 25 MHz, < 75 mV; 25 MHz to max bandwidth, < 150 mV
54641A/D, 54642A/D: dc to 100 MHz, < 100 mV; 100 MHz to max bandwidth, < 200 mV

Coupling

AC (~ 3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)

Probe ID (Agilent/HP and

Auto probe sense for 54621A/622A/641A/642A

Tek compatible)

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.