Performance Characteristics

(continued)

Display System

Display

7-inch raster monochrome CRT

Throughput of Analog Channels

25 million vectors/sec per channel with 32 levels of intensity

Resolution

255 vertical by 1000 horizontal points (waveform area) 32 levels of gray scale

Controls

Waveform intensity on front panel. Vectors on/off; infinite persistence on/off 8 x 10 grid with continuous intensity control

Built-in Help System

Key-specific help in 11 languages displayed by pressing and holding key or softkey of interest

Real Time Clock

Time and date (user setable)

Measurement Features

Automatic Measurements

Measurements are continuously updated. Cursors track current measurement

Voltage (analog channels only)

Peak-to-Peak, Maximum, Minimum, Average, Amplitude, Top, Base, Overshoot, Preshoot, RMS (DC)

Time

Frequency, Period, + Width, - Width and Duty Cycle on any channels
Rise time, Fall time, X at Max (Time at max volts), X at Min (Time at min volts), Delay, and Phase on analog channels only.

Counter

Built-in 5 digit frequency counter on any channel. Counts up to 125 MHz

Threshold Definition

Variable by percent and absolute value; 10%, 50%, 90% default for time measurements

Cursors

Manually or automatically placed readout of Horizontal (X,

X, 1/

X) and Vertical (Y,

Y)

Additionally digital or analog channels can be displayed as binary or hex values

Waveform Math

One function of 1-2, 1*2, FFT, differentiate, integrate.
Source of FFT, differentiate, integrate: analog channels 1 or 2 (or 3 or 4 for 54624A), 1-2, 1+2, 1*2

FFT

Points

Fixed at 2048 points

Source of FFT

Analog channels 1 or 2 (or 3 or 4 on 54624A only), 1+2, 1-2, 1*2

Window

Rectangular, Flattop, Hanning

Noise Floor

-70 to -100 dB depending on averaging

Amplitude Display

In dBV, dBm@50

Frequency Resolution

0.097656/(time per div.)

Maximum Frequency

102.4/(time per div.)

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.