Operator's Manual

This measurement algorithm is best suited to eye diagrams that are rendered from optical rather than
electrical signals. In the presence of inter-symbol interference and/or equalization, it may not give rea-
sonable results.

The Q factor is a measure of the overall signal-to-noise ratio of the data signal. It is computed by taking
the eye amplitude (the difference between the mean values of the 1 and 0 levels) and dividing it by the
sum of the noise values (standard deviations of the 1 and 0 levels).

All of these measurements are taken in the center (usually 20%) of the eye. Users can show the Q factor
via the Measure system.

Mask Hits

Number of samples where the signal impedes the mask.

Mask Out

Number of samples where the signal does not impeded the mask.