Operator's Manual

Setting Up SDAII

Use the following steps to start your SDAII setup.

1. If you haven't done so already, touch Analysis → Serial Data on the menu bar.

2. On the Serial Data Analysis II dialog, place a checkmark in the Enable SDA box.

3. Touch the

Setup Signal Inputs

button to configure your signal input sources, crossing levels, and sig-

nal type.

Note: Touch the

Quick View

button to quickly set up and view your serial data.

4. If you are recovering clock data from the data signal (the default selection), make sure the clock

recovery input multiplexer switch is in the up position (clock recovery via the red path).

When the switch is in the up position, the Setup Ref. Clock button is disabled (grayed out).

5. Touch the

Setup Clock Recovery

button and define clock recovery settings, such as Bit Rate and PLL

settings.

OR

If you are using an explicit reference clock, click the clock recovery multiplexer switch button so that
the switch is in the down position.

6. The

Setup Reference Clock Inputs

button is then enabled. Click this button and set up your reference

clock inputs.

7. On the far right of the Serial Data Analysis II dialog, place a checkmark in the Enable Eye Meas. box

and click the

Setup Eye Measurements

button (showing its respective dialog) to define eye meas-

urement settings.

8. On the far right of the Serial Data Analysis II dialog, place a checkmark in the Enable Jitter Meas.

box and click the

Setup Jitter Measurements

button (showing its respective dialog) to define jitter

measurement settings. The Jitter Measure dialog opens displaying another flow diagram.

922968 Rev A