Applying measurements from the decode setup dialog, Applying measurements from the measure menu – Teledyne LeCroy USB2-HSIC Decoder User Manual

Page 17

Advertising
background image

Instruction Manual

l

MsgToMsg (Message to Message) - Computes the time difference from one protocol message to
another protocol message.

l

DeltaMsg (Delta Message) - Computes the time difference between two messages on a single
decoded line. 

l

Time@Msg (Time at Message) - Time from trigger to each protocol message (meeting specified
conditions). 

l

BusLoad - Computes the load of user-defined messages on the bus (as a percent).

l

MsgBitrate - Computes the bitrate of user-specified messages on decoded traces.

l

NumMessages (Number of Messages) - Computes the number of messages which match a user-
specified definition in decoded traces.

4. On the next dialog, choose a parameter location (P1-Px) in which to run the measurement.

NOTE: If you choose a location that already stores a measurement, this selection will overwrite that setup.

Applying Measurements from the Decode Setup Dialog

You can also access PROTObus MAG serial data measurements by touching the Measure button on the
Decode Setup dialog. Follow Steps 3 and 4 above to set up the measurements. Measurements are set on the
source of whichever Decoder (1-4) is currently selected on the Decode Setup dialog.

Applying Measurements from the Measure Menu

The full menu of available measurements can be accessed through the menu bar. Standard measurements
are available even if you do not have the PROTObus MAG option installed.

Choose Measure > Measure Setup and follow the usual procedures for setting up a measurement. In this case,
you will have to manually choose the source waveform to measure, as the Decoder output is not pre-selected.

922295 RevA

15