Transaction trigger, Packet trigger – Teledyne LeCroy USB2-HSIC Decoder User Manual

Page 20

Advertising
background image

USB2/USB2-HSIC Decode and Trigger

Transaction Trigger

This will trigger upon finding transactions that meet all the Token, Data, and Handshake packet criteria.
Options vary depending on the bus speed. The principal difference between a Transaction trigger and a
Packet trigger is that for a Transaction trigger, the specified packet types simply need to be found in the
transaction, whereas for a Packet trigger, the packet types and values must meet the criteria.

1. Choose Pattern Type Transaction.

2. Choose to enter values in Binary or Hexadecimal (Hex) format. The selection propagates throughout the

entire trigger setup. Toggling between formats does not result in loss of information, but will transform
the appearance of values.

3. Select the Token Type:

l

Any Type triggers on any Token packet that meets the remaining criteria.

l

In, Out, Setup, and Ping (High Speed only) trigger only when one of these Token types meet the remain-
ing criteria.

3. Enter the Token DevAddress (Device Address) and Endpoint values.

4. To add a Data criteria to the trigger, choose the Data Type, otherwise leave Any Type selected. Not Exist

will trigger when there is no Data packet in the transaction.

5. To add a Handshake criteria to the trigger, choose the Handshake Type, otherwise leave Any Type

selected. Not Exist will trigger when there is no Handshake packet in the transaction.

Packet Trigger

This type of trigger can be set on complex patterns in any packet field—token, data, or handshake—or on a 1-
16 byte user-defined data pattern. Different controls for specifying the pattern will be activated depending on
the Category selection.

T

RIGGER

O

N

A

NY

USB2 P

ACKET

Choose Pattern Type Packet and Category Any Sync.

18