Setting alarms, Taking measurements – ZOLL E Series Monitor Defibrillator Rev G NIBP User Manual

Page 10

Advertising
background image

E S

ERIES

- N

ONINVASIVE

B

LOOD

P

RESSURE

(NIBP) O

PTION

I

NSERT

10

9650-1214-01 Rev. G

2. Press the Next softkey until Auto Adjust is

highlighted.

3. Press the Inc softkey until Auto Adjust is set to OFF:

4. Press the Return softkey to return to NIBP menu.

The E Series unit repeats the selected cuff inflation
pressure for every measurement.

Setting Alarms

The E Series NIBP option includes blood pressure
alarms for the measurements shown in the following
table.

Before taking a measurement, ensure that all the NIBP
alarms settings are appropriate for the situation. See the
E Series Operator’s Guide for instructions on setting
alarms. When an NIBP or heart rate alarm is triggered,
the unit automatically initiates a single blood pressure
measurement. See the E Series Configuration Guide for
instructions on how to enable/disable this automatic
measurement.

When the E Series unit is turned on, all alarm functions
are disabled, and the alarm limits are set to their default
values. Any changes to these settings remain in effect
until either the settings are changed or 10 seconds after
the E Series unit is turned off. You can configure the
default alarm limits at power-up; see the E Series
Configuration Guide
for more information.

Setting Alarm Limits

You can change all the alarm limits using standard
E Series procedures. See the E Series Operator’s Guide
for instructions on setting alarm limits.

You can set the alarm limits using the Auto function, or
you can choose your own limits (see the preceding
table). The Auto function sets the high limit to 30 mmHg
(4.0 kPa) above and the low limit to 20 mmHg (2.7 kPa)
below the last measurement.

Enabling, Disabling, and Suspending Alarms

You can activate and deactivate all NIBP alarms, as well
as disable audible NIBP alarm indicators, using standard
E Series procedures; refer to the E Series Operator’s
Guide
for instructions.

Changing the status of one NIBP alarm automatically
sets all the other NIBP alarms to the same status. For
example, if you enable the systolic alarm, all other NIBP
alarms are automatically activated. Similarly, if you
deactivate the diastolic alarm, all other NIBP alarms are
automatically deactivated. Activating or deactivating
NIBP alarms or disabling audible NIBP alarm indicators
does not affect the status of other E Series alarms.

Suspending an NIBP alarm disables the audible
indicators until after the completion of the next
measurement. All visual alarm indicators remain active.

Triggering NIBP Measurements

You can configure the E Series unit to take a single NIBP
measurement after the heart rate alarm is triggered and/
or the NIBP alarm is triggered. See the E Series
Configuration Guide for
further details.

Taking Measurements

This section describes how to take each type of
measurement (a single measurement, a STAT
measurement, or an automatic measurement). You can
take any type of measurement whether you are in
MONITOR, DEFIB or PACER mode except in the
following cases:

The defibrillator is charged or charging.

The previous measurement occurred within
30 seconds in automatic interval measurement
mode.

To immediately abort any measurement in progress and
deflate the cuff, press the NIBP button. As a safety
feature, the E Series unit does not take measurements
within 30 seconds of another completed measurement
when in automatic interval measurement mode.

If the E Series unit finds a fault, a message displays on
the screen. See “Troubleshooting” on page 15 for a list of
NIBP display messages and their corresponding
corrective action.

Alarm
Parameter

Default
Setting

Range

(increments

of 5 for mmHg)

Systolic

High

160 mmHg
(21.3 kPa)

80-260 mmHg

(10.7-34.7 kPa)

Low

90 mmHg
(12.0 kPa)

40-140 mmHg

(5.3-18.7 kPa)

Diastolic

High

110 mmHg
(14.7 kPa)

50-200 mmHg

(6.7-26.7 kPa)

Low

50 mmHg
(6.7 kPa)

20-100 mmHg

(2.7-13.3 kPa)

Mean

High

130 mmHg
(17.3 kPa)

60-220 mmHg

(8.0-29.3 kPa)

Low

60 mmHg
(8.0 kPa)

20-120 mmHg

(2.7-16.0 kPa)



,QF



'HF



3UHY

1H[W



5HWXUQ

,QIODWLRQ3UHVVXUH

$XWR$GMXVW



PP+J



2II

Advertising