Safety considerations – ZOLL E Series Monitor Defibrillator Rev G NIBP User Manual

Page 4

Advertising
background image

E S

ERIES

- N

ONINVASIVE

B

LOOD

P

RESSURE

(NIBP) O

PTION

I

NSERT

4

9650-1214-01 Rev. G

Safety Considerations

General NIBP Safety

Read the E Series Operator’s Guide and this
manual insert before use.

Ensure that the E Series NIBP option is operated
by qualified personnel only.

Consult a physician for the proper interpretation of
pressure measurements.

Do not use on patients known to be susceptible to
bruising.

Use caution when using on elderly hypertensive
patients, as such patients may be more susceptible
to bruising.

Route patient hoses carefully to avoid patient
entanglement, strangulation or compression of
hose.

Do not select a cuff inflation pressure that exceeds
the patient’s expected systolic pressure by more
than 30-40 mmHg (4.0-5.3 kPa). The factory-
installed default cuff inflation pressure is
180 mmHg (24.0 kPa).

Do not attach the cuff to a limb being used for IV
infusion or SpO

2

monitoring. Cuff inflation might

block the infusion, causing harm to the patient or
inaccurate measurements.

Keep patient, hose and cuff as still as possible
during measurement. Patient movement or
vibrations from outside sources, particularly moving
vehicles, can degrade measurement accuracy.

Check patient regularly for signs of skin irritation or
impaired circulation in the monitored limb.

If an alarm occurs while the audible alarm
indicators are disabled, alarms do not sound;
rather only the visual alarm indicators are
displayed.

If the accuracy of measurements is suspect, first
check the patient's vital signs by alternate means.
Then check the cuff, hose and E Series NIBP
option for proper functioning.

The cuff, hose, and fitting are defibrillation-
protected. Using the NIBP option introduces no risk
for shock due to defibrillation. The cuff and hose
are non-conductive. Using the NIBP option
introduces no risk for burns due to electrosurgery.

Cuff and Hose Safety

Only use cuffs, hoses, and connectors supplied or
specified by ZOLL.

Ensure that the hose is not kinked or obstructed
before taking measurement.

Do not use damaged cuffs, hoses, or connectors.

Ensure proper cuff selection and placement to
avoid inaccurate measurements or patient injury.

Position the cuff so it is level with the heart during
measurement.

Do not place the cuff on a limb with an SpO

2

sensor

or an infusion line.

Do not repeat NIBP measurements (particularly
STAT measurements) at intervals less than
3-5 minutes over an extended period of time.
Rapidly repeating measurement can impair
circulation in the monitored limb.

Do not sterilize or immerse the cuffs or hoses.