Specifications – ZOLL E Series Monitor Defibrillator Rev G NIBP User Manual

Page 17

Advertising
background image

Specifications

9650-1214-01 Rev. G

17

Specifications

Principle of Operation

Oscillometric

Warm-up Time

Operational in less than 10 seconds

Regulatory Standards

ANSI/AAMI SP10-2002;

EN 1060-1: 1996 Specification for Non-invasive sphygmomanometers;

EN 1060-3: 1997;

EN 60601-2-30: 2000

NIBP Measurement Range

Systolic: 40 – 260 mmHg; 5.3 – 34.7 kPa

Diastolic: 20 – 200 mmHg; 2.7 – 26.7 kPa

Mean: 20 – 220 mmHg; 2.7 – 29.3 kPa

NIBP Measurement Accuracy

Per AAMI SP10 +/- 5 mmHg (0.7 kPa) mean difference;

8 mmHg (1.1 kPa) standard deviation

a

NIBP Measurement Resolution

1 mmHg; 0.13 kPa

NIBP Measurement Cycle Time

Typical: 30 seconds
Worst Case: 180 seconds

Pulse Rate Range

30 – 220 pulses per minute

Pulse Rate Resolution

1 bpm (beats per minute)

Alarm Limits

Systolic:

High 80 – 260 mmHg

High 10.7 – 34.7 kPa


Diastolic:

High 50 – 200 mmHg

High 6.7 – 26.7 kPa

Mean:

High 60 – 220 mmHg

High 8.0 – 29.3 kPa

Low 40 – 140 mmHg

Low 5.3 – 18.7 kPa


Low 20 – 100 mmHg

Low 2.7 – 13.3 kPa


Low 20 – 120 mmHg

Low 2.7 – 16.0 kPa

Operating Time (with SpO

2

, EtCO

2

and NIBP options)

A new fully charged PD 4410 bat-
tery pack at 20º C (68º F):

35 defibrillator discharges at
maximum energy (200 J), or

1.25 hours minimum of
continuous ECG monitoring and
blood pressure measurements
once every 5 minutes, or

0.75 hours of continuous ECG
monitoring/pacing at 60 mA, 70
beats per minute.

For a new, fully charged lithium ion
battery pack at 20°C (68°F):

90 defibrillator discharges at
maximum energy (200 J), or

2.75 hours minimum of
continuous ECG monitoring and
blood pressure measurements
once every 5 minutes, or

2.25 hours of continuous ECG
monitoring/pacing at 60 mA, 70
beats per minute.