ZOLL E Series Monitor Defibrillator Rev G NIBP User Manual

Page 16

Advertising
background image

E S

ERIES

- N

ONINVASIVE

B

LOOD

P

RESSURE

(NIBP) O

PTION

I

NSERT

16

9650-1214-01 Rev. G

REPEAT NIBP MEAS
alternating with

NIBP SIGNAL WEAK

There is a weak or no oscillometric signal.

Check cuff fit and positioning.

Check hose connection to unit.

Check for kinked hose.

Increase cuff inflation pressure if clini-
cally appropriate.

SYSTEM FAULT

Communication problem with the NIBP mod-
ule.

Cycle power, and retry.

If problem persists, return for service.

False high reading

Cuff too small.

Cuff not centered over brachial artery.

Cuff too loose.

Patient’s limb below level of heart.

Select larger cuff.

Readjust cuff.

Reapply cuff or select smaller cuff.

Raise patient’s limb to heart level.

False low reading

Cuff too large.

Limb above level of heart.

Select smaller cuff.

Lower patient’s limb to heart level.

Message/Symptom

Problem

User Action