Front panel removal – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 113

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

5-3

Front Panel Removal

1. Remove handle by removing the screw and washer on each

side.

2. Remove the 3 screws on the top of the panel.

3. Front panel assembly slides down. Use caution when removing

the front panel, cables run from the front panel to the System
Board. Do NOT pull the front panel far from the main chassis.

4. Gently remove black stranded cable by pulling the

connector straight out of board.