ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 234

Advertising
background image

Operator’s Shift Checklist for E Series Products (Semiautomatic)

Recommended checks and procedures to be performed at the start of each shift. For more detailed information, see
the E Series Operator’s Guide.

Date_____________________________Location __________________________Unit Serial Number ______________________

1. Condition

1st

Shift

2nd

Shift

3rd

Shift

Remarks

Unit clean, no spills, clear of objects on top, case intact

2. Multi-function Pads

1 set preconnected / 1 spare (check expiration dates)

3. Paddles (if applicable)

Paddles clean, free of gel and contaminants, not pitted

Release from paddle wells easily

4. Inspect cables for cracks, frays, broken wires, connector pins

A. ECG electrode cable, connector

B. Defibrillator paddle cables, paddle handle

C. Multi-function cable, connector

5. Batteries

A. Fully charged battery in unit

B. Fully charged spare battery available

6. Disposable supplies

A. Electrode gel or gel patches

B. MFE Pads in sealed pouches — 2 sets

C. ECG electrodes

D. Recorder paper

E. Alcohol wipes

F. Razors

7. Memory Card

Empty memory card install in unit (if applicable)

8. Operational checks

A. Power On Sequence

Turn unit to ON, 4-beep tone heard

B. Defibrillator

Multi-function cable connected to test connector: CHECK PADS displays.

Press ANALYZE button, unit charges to 30 J

Press and hold SHOCK button, TEST OK is displayed and printed

Attach MFC to ECG Simulator, set to VF

Verify CHECK PATIENT prompt is displayed

Press ANALYZE. Verify unit charges to 120 J

Press SHOCK, verify shock was delivered.

C. Paddles (if applicable) (Manual mode only)

Paddles in holders: Set defib energy level to 30 joules, and simultaneously
press and hold both defib discharge buttons; TEST OK message on recorder.

D. Pacer Operation (Manual mode only)

Multi-function cable not connected to Test Connector

Turn to PACER, set pacer rate to 150 ppm, press RECORDER button

Pacer pulses occur ever 2 large divisions (10 small divisions)

Press 4:1 button, pulses occur every 8 large divisions

Set PACER OUTPUT to 0 mA, no CHECK PADS prompt

Set PACER OUTPUT to 16 mA, CHECK PADS prompt and alarm

Reconnect Multi-function cable to test connector.

Press Clear Pace Alarm softkey; CHECK PADS prompt disappears and Pace
alarm stops.

E. Recorder

Press RECORDER button; recorder runs. Press again; recorder stops.

Inspect recorder printout

9. Please check the appropriate box after each use of this checklist.

Signatures

No action required

1st ___________________

2nd __________________

3rd ___________________

Minor problem(s) corrected
Disposable supplies replaced
Major problem(s) identified — UNIT OUT OF SERVICE