E Series Service Manual
5-6
9650-1450-01 Rev. D
3. 2 Cutout Style: Remove black rivet by pulling it straight up.
Then remove plastic isolator.
3A. Original Style: Remove 2 screws, and lift out shield.
4. Remove 3 cables from the control board.
5. Remove display by lifting from the bottom.
Note:
During reassembly, sit display in the bottom first and then push
down on top.