Cleaning the capnostat 5 co, Sensor, Cleaning reusable airway adapters – ZOLL R Series Monitor Defibrillator Rev A ETCO2 User Manual

Page 10

Advertising
background image

R S

ERIES

— E

ND

-T

IDAL

C

ARBON

D

IOXIDE

(E

T

CO2) O

PTION

I

NSERT

8

www.zoll.com

9650-0915-01 Rev. A

b.

Press the sensor and airway adapter together until they click (see Figure 3).

3.

If the unit displays the CHECK CO2 ADAPTER message, follow steps a through c, then go
to step 4.

a. Verify proper connection of the adapter to the sensor.

b.

Verify that the airway adapter windows are clean and dry.

c.

If the adapter is properly connected, and the windows are clean and dry, then zero the

adapter as described in “Zeroing the Mainstream CAPNOSTAT 5 CO2 Sensor/Airway
Adapter” on page 6.

4.

Ensure the patient seals his or her mouth completely around the mouthpiece and breathes
normally.

A nose clip may be needed if the patient is exhaling through the nose. It is important that all, or
most, of the exhalation be routed through the airway adapter.

Cleaning the CAPNOSTAT 5 CO

2

Sensor

The outside of the sensor or module may be cleaned and disinfected by wiping with 70%
isopropyl alcohol, a 10% bleach solution, or mild soap.

After cleaning, wipe with a clean, water-dampened cloth to rinse. Dry before use.

Cleaning Reusable Airway Adapters

Reusable airway adapters may be cleaned by rinsing in a warm soapy solution, followed by
soaking in a liquid disinfectant such as 70% isopropyl alcohol, a 10% bleach solution, Cidex

®

or System 1

®

(refer to the disinfectant manufacturer’s instructions for use). Adapters should

then be rinsed with sterile water and dried.

The adult reusable airway adapter may be autoclaved at 121° C (250°F) for 20 minutes,
unwrapped.

Note:

The Neonatal reusable airway adapter cannot be autoclaved.

Before reusing the adapter, ensure the windows are dry and residue-free, and that the adapter
has not been damaged during handling or by the cleaning process.

Figure 3