Specifications

This section summarizes the specifications of the R Series End Tidal Carbon Dioxide (EtCO

2

)

option.

CAPNOSTAT 5® CO

2

Sensor

Transducer Type

Mainstream

Principle of Operation

Non-Dispersive Infrared (NDIR) single beam optics,
dual wavelength, no moving parts.

Warm Up Time

Full specifications within 2 minutes at an ambient
temperature of 25º C. Capnogram in 15 seconds.

EtCO

2

Measurement Range

0 - 150 mmHg
0 - 20%
0 - 20 kPa

EtCO

2

Accuracy

(at 760 mmHg, ambient
temperature of 25°C)

0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
101 - 150 mmHg, ±10% of reading

(At respiration rates > 80 breaths per minute, all
ranges are ±12% of actual.)

EtCO

2

Resolution

0.1 mmHg 0 - 69 mmHg
0.25 mmHg 70 - 150 mmHg

EtCO

2

Stability

Short-Term Drift: Drift over four hours < 0.8 mmHg.

Long-Term Drift: Accuracy specification will be
maintained over a 120 hour period after zeroing.

EtCO

2

Noise

RMS noise of the sensor < 0.25 mmHg at
7.5% CO

2

.

EtCO

2

Rise Time (10-90%)

< 60 ms (Adult/pediatric adapters)

< 60 ms (Infant/pediatric adapters)

Respiration Rate Range

2 - 150 breaths per minute

Respiration Rate Accuracy

±1 breath per minute

Sample Flow Rate

N/A

Compensations

Barometric pressure 400 - 850 mmHg (automatic).
Operator selectable O

2

/N

2

O compensation.

EtCO

2

Alarm Limits

User selectable, High 5 - 100 mmHg, Low
0 - 95 mmHg, OFF.

Respiration Rate (RR) Alarm Limits

User selectable, High 5 - 150 respirations per
minute, Low 0 - 100 respirations per minute, OFF.

Halogenated Agents

Specification allows for halogenated anesthetic
agents that may be present at normal clinical levels.
The presence of Desflurane in the exhaled breath
beyond normal values (5-6%) may positively bias
carbon dioxide values by up to an additional 2-3
mmHg.

Airway Adapter Deadspace

Adult < 5 cc
Infant < 1.0 cc