Physiological Monitoring

The physiological monitoring menu includes the following softkeys: Options, Param, Code
Marker
, Report Data, and Alarms.

Param Softkey

When you press the Param softkey, the following softkeys are displayed: ECG, SpO2, NIBP,
CO2, and Return.

Note:

SpO2, NIBP, and CO2 will only appear if these options are installed on your unit.

Press the Return softkey to return to the physiological monitoring menu.

Selecting the CO2 parameter causes the following softkeys to appear: Zero, Average, Comp.,
Zoom, Disable EtCO2 (or Enable EtCO2), and Return.

The Zoom softkey will only appear if a capnogram is currently displayed.

Pressing the Disable EtCO2 softkey puts the sensor into the power saving sleep mode.
Pressing Enable EtCO2 softkey takes the sensor out of sleep mode and turns the heater on for
normal operation.

Press the Return softkey to return to the physiological monitoring menu.

Zero Softkey

Adapter zeroing should be performed whenever you switch between reusable and disposable
adapters, or if a CAPNOSTAT 5 CO

2

sensor is being connected to the unit for the first time.

Zeroing may be necessary if the unit displays the message ZERO CO2 ADAPTER.

Note:

Do not attempt zeroing for 20 seconds after removing the adapter from the patient’s
airway. This time allows any CO

2

remaining in the adapter to dissipate before zeroing.

Do not attempt to zero while the adapter is in the patient’s airway. Zeroing with CO

2

in

the adapter can lead to inaccurate measurements and/or other error conditions. If you
attempt zeroing while CO

2

remains in the adapter, the time required to zero may be

increased. If zeroing cannot be completed, the message ZERO FAILED will be
displayed. If this occurs, clear any occlusion in the adapter, remove any source of CO

2

,

wait 20 seconds, and try zeroing again.

Press the Zero softkey to initiate adapter zeroing.

The unit displays the ZEROING CO2 ADAPTER or ZEROING CO2 MODULE message during
the zeroing process, which is typically finished in 15-20 seconds.

The unit displays the ZERO DONE message when the zeroing process is complete.

SpO2

ECG

Return

CO2

NIBP

Average

Zero

Return

Zoom

Comp.

Disable

EtCO2