Check out procedures, Mainstream etco, Capnostat 5 co – ZOLL R Series Monitor Defibrillator Rev A ETCO2 User Manual

Page 15: Sensor)

Advertising
background image

Check Out Procedures

9650-0915-01 Rev. A

R Series End-Tidal Carbon Dioxide (EtCO

2

)

13

EtCO

2

value

Respiration rate

The recorder runs continuously until the button is pressed again. If selected, the capnogram is
also printed at a fixed scale of 40 mmHg/cm or 5.3 kPa/cm. All waveforms printed by the
recorder are delayed by 6 seconds relative to their occurrence.

Check Out Procedures

The following procedures verify that the EtCO

2

option is functioning properly.

Mainstream EtCO

2

(CAPNOSTAT 5 CO

2

Sensor)

1. Connect the CAPNOSTAT 5 CO

2

sensor cable to the yellow EtCO

2

connector at the back of

the R Series unit.

2.

Connect an airway adapter to the CAPNOSTAT 5 CO

2

sensor.

3.

Turn the selector switch to MONITOR mode (ON for BLS/Plus units and select Manual
mode).
If the CO

2

box displays DISABLED, enable the sensor by pressing the Param softkey, then

the CO2 softkey, then the Enable EtCO2 softkey.

4.

Wait for the CO

2

sensor to warm up. The message CO2 WARM UP is displayed for

approximately one minute.

5.

Perform a zero procedure if necessary (see “Zeroing the Mainstream CAPNOSTAT 5 CO2
Sensor/Airway Adapter” on page 6).

6.

Breath normally into the adapter.

7.

Verify that the unit displays appropriate readings in the EtCO

2

display area of the monitor.

8.

With alarms enabled, verify that the patient alarms are functional by adjusting the high and
low limits until the unit:

Emits a continuous audio tone.

Highlights the alarming parameter’s value and flashes the alarm symbol on the display.