8. PARAMETER LISTS

T0802-5.EPS

Write

Mode

Explanation

Factory

Default

Relative

Index

Parameter Name

Index

Block Alarm status is shown by this parameter.
Valid range is bit 0: Discrete alm, bit7: Block alarm
(only bit 0 and bit 7 are effective).
Initial value is

Current: 0, Unacknowledged: 0, Unreported: 0,
Disable: 0X0000 (all alarms are enable).

Selects whether or not the adhesion diagnostic function will be
carried out. (If the judgment value for Level 3 exceeds, a warning
is displayed; and if the value for level 4 exceeds, an alarm is
displayed. Valid range is

1: No, 2: Yes

This parameter is used in order to set the resistance value for
judgment of Level 1. Valid range is 0.0 to 100.00 M Ohm

This parameter is used in order to set the resistance value for
judgment of Level 2.Valid range is 0.0 to 100.00 M Ohm

This parameter is used in order to set the resistance value for
judgment of Level 3.Valid range is 0.0 to 100.00 M Ohm

This parameter is used in order to set the resistance value for
judgment of Level 4.Valid range is 0.0 to 100.00 M Ohm

This parameter displays the value measured using the adhesion
diagnostic function. Valid range is 0.0 to 1000.00 M Ohm

85 2085

ALARM_SUM

86 2086

ADHESION_CHECK

1:No

O/S

87 2087

ADHESION_LEVEL1

0.10

O/S

88 2088

ADHESION_LEVEL2

0.50

O/S

89 2089

ADHESION_LEVEL3

1.00

O/S

90 2090

ADHESION_LEVEL4

3.00

O/S

91 2091

-

ADH_MEASURE_
VALUE