12 switching to deep sleep mode, 13 switching to silence mode, Switching to deep sleep mode -22 – Yokogawa EJX530B User Manual

Page 55: Switching to silence mode -22, Caution

Advertising
background image

<8. Setting Parameters>

8-22

IM 01C27F01-01EN

8.3.12 Switching to Deep Sleep Mode

When the instrument will not be used for a long

time, switch the instrument to deep sleep mode to

conserve battery power.

To switch to deep sleep mode, follow the procedure

below.

• Procedure to call up the switch-to-deep-sleep

parameter

TRANSDUCER block: Special Cmd

Set Deep Sleep mode (Standby) to the Special

Cmd parameter for the TRANSDUCER

block. To start from deep sleep mode, either

remove and insert the battery pack, or use

the provisioning device tool or the device

configuration tool via infrared communication.

CAUTION

After setting the deep sleep mode by infrared

device configuration tool, keep the infrared port

of device away from any other infrared signals.

NOTE

• Transmitter becomes the stop state after

setting deep sleep mode and cannot reply

any request from the device configuration

tool via wireless communication.

• For this reason, there is the case that an

error is display on the device configuration

tool via wireless communication.

• To wake up from deep sleep mode, please

pull battery pack, and wait more than 30

seconds before attaching battery back.

8.3.13 Switching to Silence Mode

This is a function to pause the instrument when

it cannot join the field wireless network after a

specified time has elapsed. This function is effective

in conserving battery power when, for example, the

installation of the field wireless integrated gateway

is delayed compared to that of field wireless

devices. The default value is 28800 seconds (8

hours). Thereafter, a cycle of a one-hour pause and

six-minute search is repeated until the instrument

can join the field wireless network.

• Procedure to call up the switch-to-silence

parameter (Radio Silence)

UAPMO block: Radio Silence

Set 0 to 2

31

seconds for the Radio Silence

parameter of the UAPMO block. If 0 is set, the

Radio Silence parameter is invalid. To start

from silence mode, either remove and insert

the battery pack, or use the provisioning device

tool or the device configuration tool via infrared