<Appendix B Messages Summary>

B-2

IM 11B03G03-03E

n

Alarm Windows

Message

Description

Corrective Measures

History

deletion

Alarm history cannot be

cleared in User level A.

Invalid attempt to reset the alarm

history at the wrong user level.

Alarm histories can be reset at

user level B or C only.

Creation of

user alarm

description

Factory alarm

descriptions cannot be

changed. Select User

defined alarm.

A number other than the user

defined alarm numbers has been

specified when creating a user

alarm description.

Confirm the valid alarm numbers,

and then select a correct alarm

number.

n

Analysis Result Window

Message

Description

Corrective Measures

Memory

Memory AI location Error.

Not enough memory.

The memory is too low.

Check that memory is reserved

sufficiently for work.

File

operation

File name is incorrect.

This file is not XXX.

Invalid attempt to open the file

that does not contain the data

corresponding to the onscreen

window.

Open the file appropriate for the

on-screen window.

Read error. Data is

incorrect. Error occurred

reading a file.

Failed to open the file because it

contained invalid data.

Open the file that contains the

valid data.

Error occurred while

saving file. File cannot be

saved.

An error occurred while saving

file. Filed to save the file.

Eliminate the cause of the error

and then retry saving the file.

Data has been changed.

Do you want to save it?

Changes have been made to the

data of the window.

If the file needs to be saved

before the window is closed,

select “Yes.”

Data

editing

Entered valve is incorrect.

Only a numerical value

can be entered here.

Invalid attempt to type in a data

value other than numerals.

Type in a numeric data value.

Automatic

saving

Entered value is incorrect.

Automatic saving cycle

must be between 1 and

250.

An invalid number has been

typed in as the interval for

automatic saving.

For the interval for automatic

saving, type a whole number

from 1 to 250.

Graph

creation

Graph cannot be drawn.

No data available.

Cannot create a graph because

the currently open window does

not contain graph data.

Create a graph when data are

available.

Graph cannot be drawn.

Data is incorrect.

Cannot create a graph because

the selected range for the graph

being created contains invalid

data. This message also appears

if the selected range for the

graph does not contain any data.

Once again check the data for

the selected range of graph. Also

check that the range contains

necessary data.

Selection error.

Necessary data is not

selected.

Some of the data necessary

to create the graph remain

unselected.

Once again check that none

of the necessary data remain

unselected.

Selection error. You can

select 1 to 6 Peak names.

Seven or more peaks have been

selected at one time during peak

selection.

A graph can contain no more

than six peaks at a time. Select

no more than six peaks.

Selection error. Start time

is set after End time.

The starting time has been set at

an later time than the ending time

during time-axis setting (manual).

Set the starting time at a time

earlier than the ending time.

Graph cannot be drawn.

Min./max. axis scale of a

Logarithm graph cannot

be set or under zero.

The upper or lower limit of

the scale has been set at a

value smaller than zero during

the setting (manual) of the

logarithmic axis.

A logarithmic axis cannot indicate

values smaller than zero.

When setting the upper and

lower limits of a logarithmic axis,

use values larger than zero.

4th Edition : Mar. 21, 2012-00