Function of Buttons and Available Information

The following table summarizes the function of each button and the available information.

Button

Function and Available Information

Status (command)

button

Indicates the current mode of the analyzer. By clicking on the button of the desired

status, the status can be changed.

Operation mode

(command) button

Indicates the current operation mode and the status in which a mode change is

commanded and reserved. By clicking on the button of the desired operation mode,

the mode can be changed. "Cancel" command button is used to cancel the reserved

operation mode.

Measuring status

button

Only used to command the measureing status by clicking on the button of the desired

measuring status. "Cal (Val) start" and "Cal (Val) end" buttons are used to start/end

the calibration/validation manually. "Cancel" command button is used to cancel the

scheduled measurement. "Cal (Val) mathod" button is used to change the calibration/

validation method.

Range change

button

Used to change the range.

Help indicate button

Indicates hepl of GCMT.

Measuring state

Indicates the current measuring state.

Next measuring

Indicates the measuring state scheduled next, if the measuring state has been

changed.

Method

Indicates the current method number.

HELP !

To be able to work with the operation status ribbon, the user level must be changed.

See Also

For details on how to change the user level, see Section 3.4.

n

Chromatogram Display Area

This area displays the latest chromatogram(s) and the temperature pattern of the programmed-

temperature oven (the dual oven) or of the isothermal oven (the single oven) by the analysis

cycle. If there are two detectors, the area shows two chromatograms.
Double-clicking in this area opens the Chromatogram window.

See Also

For details on the Chromatogram window, see Chapter 4.

l

Area Layout

F0308.ai

Chromatogram

Temperature

The time elapsed up to now is shown as a broken line.

4th Edition : Mar. 21, 2012-00

Advertising