<3. Analyzer Operation Window>

3-20

IM 11B03G03-03E

n

Changing the Peak Individual Setting

(1) Click on the [Peak] command in the [Set Up] menu. Click on the command of the information

you want to change, such as [Name] and [Gate Std/On/Off Time] commands.

(2) A dialog box (e.g., Name dialog box) pops up.

F0333.ai

Specify the stream number and peak number you want to change. The information of the

current state is displayed.

(3) Change the information in the dialog box, and then click on the [Execute] button. A

confirmation dialog box pops up.

F0334.ai

(4) Click on the [Yes] button.
(5) A message appears to inform you that the data has been changed successfully.

F0335.ai

(6) Click on the [OK] button to finish.

3.9

Uploading and Downloading the Parameter

Settings

When duplicating the settings of the Parameter List Setup window on the LCD panel to another

analyzer or when making a backup of the settings to re-configure them, use the operations of

uploading/downloading parameters.

NOTE

Parameter settings can be uploaded or downloaded only at user level C and in the Stop mode in

either Process or Lab status.

4th Edition : Mar. 21, 2012-00