Installing the maintenance terminal, Nsystem configuration, Lsoftware conditions – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 8: Lhardware conditions

Advertising
background image

<Installing the Maintenance Terminal>

1

IM 11B03G03-03E

Installing the Maintenance Terminal

This chapter describes the installation of the Maintenance Terminal (GCMT) in the

personal computer system being used. This installation assumes that your PC system

is already in the ready state under the following conditions and also that you have some

knowledge of how to use the system.

n

System Configuration

l

Software conditions

Check that the software meets the following conditions:

Microsoft’s Windows 7 Professional 32bit Service Pack 1 or

Windows Vista Business Edition or

Windows XP Professional Service Pack 3

l

Hardware conditions

Before installing GCMT, check that the hardware meets the following conditions:

• Models supported: IBM PC or computable machine that can run Microsoft Windows and for

which Ethernet is available

OS

Windows 7, Windows Vista

Windows XP

CPU

1 GHz or better

Pentium II 350 MHz or better

HDD

15 GB or more for OS

10 GB or more for application

128 MB or more

RAM

At least 1 GB

At least 20 MB

Display

680 x 480 VGA or higher,

and be viewable with equal to or more than 256 colors

• For the hard disk, a data storage capacity should be secured separately according to your

PC system format, in addition to the capacity for the program.

• Communication function: Make sure that the standard serial communication port is

connected to the analyzer via the communication converter (Model K9404LA provided with

a feature for maintaining the explosion-proof capabilities)-see the figure below.

K9404LA

GCMT

Personal computer

Software package

Field

Instrument room

GC1000

Mark II

analyzer

RS-422

(up to 1 km)

RS-232C

(up to 15 m)

Communication

converter

Serial

communication

Serial

communication

Power-off signal

RS-422/RS-232C converter provided

with a feature for maintaining

explosion-proof capabilities

Installation

F01.ai

4th Edition : Mar. 21, 2012-00

Advertising