<7. HYDROGEN PURITY / PARAMETER SETTING>

7-4

IM 11T3E1-01E

7.1.4

Setting Parameters at Service Level

Press the [*] key and enter the password (XXX) to select the service level. When you select this

level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the

[ENT] key.

Table 7.4 Setting Parameters at Service Level

Code

No.

Item

Display

Data to be Set

(or Conditions)

Remarks

01 Setting of hold during maintenance

Hold enable/disable

Setting of preset hold value

*M_HLD

*PR.SET

· Disable : 0

· Enable (value immediately

before) : 1

· Enable (preset value) : 2

-10.0 to 110.0

See Figure 7.8 for key operation.

If “preset value” is selected in the

preceding parameter.

02 Setting of hold against errors

Hold enable/disable

Setting of preset hold value

*E_HLD

*PR.SET

· Disable : 0

· Enable (value immediately

before) : 1

· Enable (preset value) : 2

-10.0 to 110.0

See Figure 7.9 for key operation.

If “preset value” is selected in the

preceding parameter.

03 Setting of hydrogen purity range

Hold enable/disable

Setting of the preset hold value

*H_HLD

*PR.SET

· Disable : 0

· Enable (value immediately

before) : 1

· Enable (preset value) : 2

-10.0 to 110.0

See Figure 7.10 for key operation.

If “preset value” is selected in the

preceding parameter.

04 Setting of output-smoothing constants

*SMOTH

00 to 60

See Figure 7.11 for key operation.

05 Setting of contact outputs status

*CNTCT

00 to 15

See Figure 7.12 for key operation.

10 Setting of pressure compensation

Enable/disable/fixed value

Setting of fixed value for pressure

correction

*P.COMP

*P.FIX

· Disable: 0

· Enable (measured value) : 1

· Enable (fixed value) : 2

0.10 to 999.99kPa, 0.0001 to

9.9999MPa, 0.015 to 99.999 psi

See Figure 7.13 for key operation.

If “fixed value” is selected in the

preceding parameter.

11 Setting for measurement compen-

sated density

Setting of reference temperature

Setting of reference pressure

*C.D.TMP

*C.D.PRS

-20.0 to 80.0 °C, -4.0 to 176.0 °F

0.10 to 999.99kPa, 0.0001 to

9.9999MPa, 0.015 to 99.999 psi

See Figure 7.14 for key operation.

12 Setting of gas pressure

Setting of zero-gas pressure

Setting of span-gas pressure

*Z_PRS

*S_PRS

0.01 to 999.99kPa, 0.0001 to

9.9999MPa, 0.015 to 99.999 psi

0.10 to 999.99kPa, 0.0001 to

9.9999MPa, 0.015 to 99.999 psi

See Figure 7.15 for key operation.

14 Setting for remote semi-automatic

calibration

*REMOT

· Disable : 0

· Enable : 1

Open contact : AIR in CO

2

Closed contact : H

2

in CO

2

See Figure 7.16 for key operation.

20 Selection of pressure unit

*PRES.U

· kPa : 0

· MPa : 1

· psi : 2

See Figure 7.17 for key operation.

21 Selection of density unit

*DENS.U

· kg/m

3

: 0

· lb/ft

3

: 1

See Figure 7.18 for key operation.

23 Selection of temperature unit

*TEMP.U

· °C : 0

· °F : 1

See Figure 7.19 for key operation.

31 Showing/hiding of negative measured

values (-)

*MINUS

· Show : 0

· Hide : 1

See Figure 7.20 for key operation.