<10. HYDROGEN PURITY / CALIBRATION PROCEDURE>

10-1

IM 11T3E1-01E

10. HYDROGEN PURITY /CALIBRATION

PROCEDURE

This chapter describes how to calibrate of Hydrogen purity meter.

If select Density meter, see Chapter 8.

If select Calorie meter, see Chapter 9.
Zero and span calibration can be carried out using standard gases (zero and span gases). The

analyzer uses hydrogen (100%) as the zero gas and carbon dioxide (100%) as the span gas.

The analyzer is designed so that calibration is done manually. The basic calibration procedures

and operations are as follows.

10.1 Basic Calibration Procedure

Operation level

Manual calibration mode

See the Process in Figure 7.2.

See the Process in Figure 10.2.

F1001.ai

10.2 Preparation for Calibration

Prepare for calibration during the initial operation of the analyzer by entering manual calibration

at the operation level.

[NO]

*SERVC

[NO]

[NO]

*RANGE

[NO]

*ALARM

*CAL.DT

X

*C_HLD

XXX.X

*PR.SET

F1002.ai

[YES]

*

3

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*3: This step is skipped if the parameter [*C_HLD] is set to a value other than "2".

*

1

[ENT]

[ENT]

*

2

(Setting of output hold value

during calibration)

(Setting of preset value)

*

2

Figure 10.1 Calibration Parameter Setting Mode