3 setting parameters at setting level, Setting parameters at setting level -4 – Yokogawa GD40 Gas Density Detector User Manual

Page 93

Advertising
background image

<6. CALORIE / PARAMETER SETTING>

6-4

IM 11T3E1-01E

6.1.3

Setting Parameters at Setting Level

Press the [*] key and enter the password (XXX) to gain access to this level (see Figure 6.5 for key

operation).

Table 6.3 Setting Parameters at Setting Level

Mode/Setting parameter

Display

Data to be Set

(or Conditions)

Remarks

Analog output setting

Output 1

Set a parameter to be output

Set the zero and span points

Physical density

Compensated density

Calorific value

Output 2 (same as output 1)

*RANGE

*OUT1

*DENS

*C_DNS

*CALRY

*TEMP

*PRESS

*Z_DNS,

*S_DNS

*Z_CP.D,

*S_CP.D

*Z_CAL,

*S_CAL

*OUT2

Physical density

Compensated density

Calorific value

Gas temperature

Gas pressure

00.0000 to 60.0000 kg/m

3

0.00000 to 4.00000 lb/ft

3

0.0000 to 6.0000 kg/m

3

0.00000 to 0.40000 lb/ft

3

000.000 to 133.000 MJ/m

3

0.0000 to 3.5000 kBTU/ft

3

See Figure 6.5 for key operation.

No setting is required if the parameter is

temperature or pressure.

Setting calibration data

Zero-point calorific value

Zero-point density

Span-point calorific value

Span-point density

Output hold value during calibra-

tion

Setting of preset value

(Only if the preceding parameter

is set to 2 (preset value))

*CAL.DT

*Z_CAL

*Z_DNS

*S_CAL

*S_DNS

*C_HLD

*PR.SET

000.000 to 133.000 MJ/m

3

0.0000 to 6.0000 kg/m

3

0.00000 to 0.40000 lb/ft

3

000.000 to 133.000 MJ/m

3

0.0000 to 3.5000 kBTU/ft

3

0.0000 to 6.0000 kg/m

3

0.00000 to 0.40000 lb/ft

3

· Disable: 0

· Enable (value

immediately before) : 1

· Enable (preset value) : 2

-10.0 to 110.0

See Figure 9.2 for key operation.

Alarm-point setting

Select a parameter for which an

alarm is set

Physical density

Compensated density

Calorific value

Set the high/low limits.

Physical density

Compensated density

Calorific value

*ALARM

*DENS

*C_DNS

*CALRY

*L_DNS,

*H_DNS

*L_CP.D,

*H_CP.D

*L_CAL,

*H_CAL

00.0000 to 60.0000 kg/m

3

0.0000 to 6.0000 kg/m

3

000.000 to 133.000 MJ/m

3

See Figure 6.6 for key operation.