<1. Specifications>

1-5

IM 11T3E1-01E

1.1.2

GD40G, T, V, R Detector

l

Material exposed to gas

SUS 316 stainless steel, Acrylonitrile Butadiene Rubber and Fluorine-contained Rubber

(O-ring)

l

Ambient conditions

Temperature:

-10 to 60 °C (14 to 140 °F)

Humidity:

5 to 95 %RH

Installation:

Pipe-mounted or on panel

Construction:

Explosion-proof and Intrinsically safe. (See 1.1.1 “System Components”)

Though the detector construction makes it relatively insensitive to sudden changes in

the gas temperature, extra precision can be achieved by keeping ambient tempera-

ture conditions as constant as possible. In measurements where optimum precision is

required it is therefore not recommended to install the detector in an outdoors environ-

ment, especially not if such installation is prone to direct sunlight.

l

Finish

Cover:

Equivalent to Munsell 0.6GY3.1/2.0

Case:

Equivalent to Munsell 2.5Y8.4/1.2

l

Weight

Approximately 7 kg (with pipe mounting bracket)

l

Detector unit

When the system is ordered to be used as a hydrogen purity meter, an optional pressure

transmitter is required for pressure compensation.

• If /EJAJ1, /EJAF2, /EJAF3 or, /EJAF4 are ordered, the detector unit and the

pressure transmitter and the tubing in between will all be integrated on a single mounting

plate. This allows the space where the pressure transmitter is normally mounted to be used

effectively for other purposes.

1.1.3

GD402G, T, V, R Converter

l

Display

Reading:

Digital (6 digits maximum; resolution: 0.0001 kg/m

3

)

Data items shown:

Measured value:

Always on display.

Alarm indications: Abnormal concentration, abnormal pressure range of input and abnormal

value of calibration.

Parameters for calibration: Time of calibration, settling time, starting time of calibration and

calibration cycle

Self-diagnostic indications: Sensor oscillation shutdown, abnormal oscillation frequency of

sensor, failure in sensor temperature detection, failure in A/D conversion

stage and memory failure.

Alarm settings:

The contact state can be set to either “normally open (NO)” or “normally

closed (NC)” depending on the needs of the application.

Temperature:

Temperature of gas being measured