Series IN 500 Operating Manual

Controls and Connections 16

3.2 Connection cable

The connecting cable has to be selected according to
the following criteria:

Shielded
Diameter 3 - 6.5 mm
Number of wires: 2 to 12 (as required)
- 2 for power supply.

additionally, if required:
- 1 for analog output
- 1 for sensor head temperature

- 1 GND for analogue output and /
or sensor head temperature
- 3 or 4 for digital interface
- 2 for relay contact

3.3 Connection

1. Prepare the cable as shown in Fig. 1 .

2. The shielding of the connecting cable has to

be pushed over the plastic sleeve
2 mm further than the O-ring.

3. Fix the cable with the clamp screw.
4. At state of delivery, an internal connection

between contact 11 and the housing exists.
Should the housing have its own earth
potential caused by an electric connection to
an object, the internal connection has to be
removed. Fig. 2 .

The screw bars inside the converter can be lifted up

for easy installation of the wires (input/output
cables and sensor head) Fig. 3 .

_________________________________________________

Caution: Before connecting the pyrometer

to other instruments (e.g. controller, PLC,

etc.), the correct analog output (current, voltage or
thermocouple) corresponding to the input of the
instrument, has to be adjusted at the converter. A

wrong output might damage the connected
instrument.

Fig. 3

Fig. 2

Grounding Wire Bridge