5 measurement online trend – LumaSense Technologies IGA 320/23-LO User Manual

Page 27

Advertising
background image

IGA 320/23-LO Operating Manual

Software InfraWin 27

In the lower part of the window the connection with the preset baud rate can be checked. Here
the command was sent 500 times with 19200 baud. It has taken 4.011 seconds without
transmission errors.
Emi: AutoFind: If the true temperature of the measured object is known, you can calculate the
emissivity of the measured object using the Emi: AutoFind function:
A measured temperature is displayed with the current set
emissivity (in this example 100%) (here: 219.0°C).

If you press Emi: Autofind, a window will open that allows

you to enter the “true” temperature.

Once the temperature entry has been entered and
confirmed with Calculate, InfraWin will then calculate the

emissivity which occurs with the new temperature. This is
displayed immediately and can be used for further
temperature measurement.

5.5 Measurement online trend

The measurement
function allows

you to access a number of input
tabs located on the left side of the

screen. The main or home tab is the
Output Screen. You can toggle the
input tabs on and off by clicking
them. This window displays:

temperature as graphical

diagram

internal temperature of the

instrument

current temperature

quantity of the measured values and file size of the current measurement

The example shows a sample reading over the period of approximately 13 seconds with a

temperature range between 150 and 1200 °C. The final temperature (at the end of the reading)
is 149 °C.

The Mark Zone button allows you to color mark a temperature range for easier recognition.
The Threshold button allows you to set a temperature value as a baseline to prevent recording

values above or below the baseline temperature. This allows you to keep the output file size
small.

The Scaling Trend button allows you to scale temperature trend view.

Advertising